Effects of Charge Carrier Trapping on Image Resolution of Multilayer Photoconductive Detectors: Application to Amorphous Selenium X-ray Detectors
The effects of charge carrier trapping on image resolution in multilayer photoconductive imaging detectors was analyzed by developing an analytical model for calculating the modulation transfer function (MTF) of the imaging detectors. The MTF model was developed by considering the charge carrier tra...