Half-threaded holes markedly increase the fatigue life of locking plates without compromising screw stability
Aims: To determine whether half-threaded screw holes in a new titanium locking plate design can substantially decrease the notch effects of the threads and increase the plate fatigue life. Methods: Three types (I to III) of titanium locking plates were fabricated to simulate plates used in the femur...
Main Authors: | , , |
---|---|
Format: | Article |
Language: | English |
Published: |
The British Editorial Society of Bone & Joint Surgery
2020-10-01
|
Series: | Bone & Joint Research |
Subjects: | |
Online Access: | https://online.boneandjoint.org.uk/doi/epdf/10.1302/2046-3758.910.BJR-2019-0237.R2 |