Less Time Is Less Motion: Analysis of Practical Efficiencies Gained With a Modified Workflow Integrating Planar kV Midimaging With CBCT for Spine Stereotactic Body Radiation Therapy

Purpose: Our purpose was to optimize an image guided radiation therapy (IGRT) workflow to achieve practical setup accuracy in spine stereotactic body radiation therapy (SBRT). We assessed the time-saving efficiencies gained from incorporating planar kV midimaging as a surrogate for cone beam compute...

Full description

Bibliographic Details
Main Authors: David Y. Hu, PhD, Yiwen Xu, PhD, Yu-Hui Chen, PhD, Marjan Khosravi, MS, Yulia Lyatskaya, PhD, Jeremy S. Bredfeldt, PhD, Fred L. Hacker, PhD, Tracy A. Balboni, MD, Alexander Spektor, MD, PhD, Daniel Cagney, MD, Raymond Mak, MD, Mai Anh Huynh, MD, PhD
Format: Article
Language:English
Published: Elsevier 2022-09-01
Series:Advances in Radiation Oncology
Online Access:http://www.sciencedirect.com/science/article/pii/S2452109422000689