Risk Assessment-Oriented Design of a Needle Insertion Robotic System for Non-Resectable Liver Tumors
Medical robotics is a highly challenging and rewarding field of research, especially in the development of minimally invasive solutions for the treatment of the worldwide leading cause of death, cancer. The aim of the paper is to provide a design methodology for the development of a safe and efficie...
Main Authors: | Bogdan Gherman, Nadim Al Hajjar, Paul Tucan, Corina Radu, Calin Vaida, Emil Mois, Alin Burz, Doina Pisla |
---|---|
Format: | Article |
Language: | English |
Published: |
MDPI AG
2022-02-01
|
Series: | Healthcare |
Subjects: | |
Online Access: | https://www.mdpi.com/2227-9032/10/2/389 |
Similar Items
-
Development of a 6-DOF Parallel Robot for Potential Single-Incision Laparoscopic Surgery Application
by: Doina Pisla, et al.
Published: (2023-10-01) -
Design and Experimental Setup of a Robotic Medical Instrument for Brachytherapy in Non-Resectable Liver Tumors
by: Paul Tucan, et al.
Published: (2022-11-01) -
Preliminary Study of a Modular MR-Compatible Robot for Image-Guided Insertion of Multiple Needles
by: Amanda M. Aleong, et al.
Published: (2022-05-01) -
Joint-Space Characterization of a Medical Parallel Robot Based on a Dual Quaternion Representation of SE(3)
by: Iosif Birlescu, et al.
Published: (2020-07-01) -
Safety Issues in the Development of an Innovative Medical Parallel Robot Used in Renal Single-Incision Laparoscopic Surgery
by: Doina Pisla, et al.
Published: (2023-07-01)