Spectrophotometric flow-injection analysis assay of tetracycline antibiotics using a dual light-emitting diode based detector

In this paper, a small dual light-emitting diode (LED)-based detector for FIA process analyser has been designed. The detector’s optical parts comprise a flow-through cell, a dual-blue LED and a photodiode. Neither mirrors nor lenses are used. The optical path for the first LED detects the blank, wh...

Full description

Bibliographic Details
Main Author: Prinya Masawat
Format: Article
Language:English
Published: Maejo University 2008-02-01
Series:Maejo International Journal of Science and Technology
Subjects:
Online Access:http://www.mijst.mju.ac.th/vol2/201-209.pdf