Fast MCsquare-Based Independent Dose Verification Platform for Pencil Beam Scanning Proton Therapy
Purpose: To commission MCsquare (a multi-cores CPU-based dose calculation engine) for pencil beam scanning (PBS) proton therapy, integrate it into RayStation treatment plan system (TPS) to create a dedicated platform for fast independent dose verification. Method: A MCsquare-based independent dose v...
Main Authors: | , , , , , , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
SAGE Publishing
2021-07-01
|
Series: | Technology in Cancer Research & Treatment |