Parallel transmit pulse design for patients with deep brain stimulation implants
Purpose Specific absorption rate (SAR) amplification around active implantable medical devices during diagnostic MRI procedures poses a potential risk for patient safety. In this study, we present a parallel transmit (pTx) strategy that can be used to safely scan patients with deep brain stimulat...
Main Authors: | , , , , , , , , , , |
---|---|
Other Authors: | |
Format: | Article |
Language: | en_US |
Published: |
Wiley Blackwell
2017
|
Online Access: | http://hdl.handle.net/1721.1/110737 https://orcid.org/0000-0001-7208-8863 https://orcid.org/0000-0002-7637-2914 |