Calibration-free measurement of liquid permittivity and conductivity using electrochemical impedance test cell with servomechanically adjustable cell constant

This paper presents a technique and a device for measuring the permittivity and conductivity of liquids and gases that does not require prior calibration by a reference sample. The technique involves precisely controlling the separation between two closely spaced spherical electrodes to make electri...