Characterization of individual submicron perfluorocarbon gas bubbles by ultrasonic backscatter

Measurements were undertaken to determine the unknown microbubble-size distribution of a dodecafluoropentane (DDFP) emulsion consisting of 1012 droplets/ml in surfactant-stabilized water. The acoustic backscatter of 2-microsecond-duration tonebursts of 30-MHz focused ultrasound was measured from the...

पूर्ण विवरण

ग्रंथसूची विवरण
मुख्य लेखकों: Everbach, E, Khismatullin, D, Flaherty, J, Roy, R
स्वरूप: Journal article
प्रकाशित: Acoustical Society of America 2005
विवरण
सारांश:Measurements were undertaken to determine the unknown microbubble-size distribution of a dodecafluoropentane (DDFP) emulsion consisting of 1012 droplets/ml in surfactant-stabilized water. The acoustic backscatter of 2-microsecond-duration tonebursts of 30-MHz focused ultrasound was measured from the emulsion as it moved in a coaxial flow. Calibration for the system was accomplished using 3-μm-radius polystyrene spheres, using a linear scattering model and literature values for polystyrene. Applying viscous linear scattering theory to the backscatter data from individual DDFP bubbles allowed inversion of the radius–backscatter relation. A mean microbubble radius of 130 nm was inferred for the DDFP population.