THE CALIBRATION OF SLOW-NEUTRON POLARIZED BEAM INSTRUMENTS

A new procedure for calibrating the efficiencies of the polarisation sensitive components in slow neutron polarised beam instruments is described. The measurement of one "flipping ratio" and two "shim ratios" allows the effective polarisation, the spin flipping efficiency, and a...