Non-contact luminescence lifetime microthermometry using scintillation sensors
The paper describes an original technique for non-contact in situ monitoring of the temperature developed for experiments in a vacuum environment at the tender X-ray beamline I23 of the Diamond Light Source. The sample temperature is established by determining the luminescence decay constant of a Bi...
প্রধান লেখক: | , , |
---|---|
বিন্যাস: | Conference item |
প্রকাশিত: |
Polish Academy of Sciences Institute of Physics
2017
|