Determination of percentage depth dose for 6 and 10 MV x-rays using Ge-doped optical fibre and TLD-100

Percentage depth doses for 6 and 10 MV x-ray beams from a linear accelerator were measured using ~1 cm long (~0.3 mg) Ge-doped optical fibre as a thermoluminescence dosimeter for two field sizes, 5 × 5 and 10 × 10 cm2. The results indicate that the Ge-doped optical fibre dosimeter is in good agreeme...

Full description

Bibliographic Details
Main Authors: Abdulla, Y.A., Amin, Y.M., Khoo, H.B.