The effect of cannulated hip screw size in the treatment of femoral neck fracture

This study aimed to investigate the mechanical performance of different cannulated hip screw size in the treatment of femoral neck fracture via finite element method. The 3D model of a proximal femur was constructed CT data of human lower limb. The femoral neck fracture was stimulated with a 40o cut...

Full description

Bibliographic Details
Main Authors: Ishak, Muhammad Farhan, Abdul Kadir, Mohammed Rafiq, Ab. Rahman, Shaifuzain
Format: Article
Language:English
Published: Penerbit UTM Press 2015
Subjects:
Online Access:http://eprints.utm.my/55618/1/MohammedRafiqAbdulKadir2015_TheEffectofCannulatedHipScrewSize.pdf